ModelConsumption, l/minMembrane area, m²PipelineWeight, kg
LF3K0.15-30.01-0.3DN1070
LF10K0.3-100.1-1.1DN1573
LF20K0.5-200.5-2.5DN20110

Product Features

Low shear

A specialized diaphragm pump designed for biological products.

Automatic TMP

Proportional control maintains transmembrane pressure.

Cassettes and hollow fibers

The membrane format is selected according to the technological task.

Weight control

Automatic weighing controls the concentration and speed of refill.

Self-draining

Emptying and in-system CIP cleaning are provided.

Process recording

Database management includes real-time logging.

Technical Parameters

Technical Parameters: Laboratory Ultrafiltration System
ParameterMeaning
Maximum working pressure5 bar
Pressure sensor-1...9 bar; accuracy ≤0.3%
pH0-14
Electrical conductivity1 µS/cm - 300 mS/cm
Contact materials316L, PP, EPDM
Operating temperature2-40 °C
Degree of protectionIP54
Power1.5 kW
Food220 V AC (±10%), 50/60 Hz

Frequently asked questions

What membrane formats are supported?

Tangential membrane cassettes and hollow fiber filter modules.

Can the system be used for lightening?

Yes. When selecting a microfiltration membrane, it is possible to remove cells and cellular fragments and clarify the liquid.

How is concentration controlled?

Automatic TMP regulation and weight control of volume and recharge are provided.