Chromatographic systems
Production chromatography system
Automated system for pilot and large scale chromatographic purification.
Product Description
Fully automated in-house developed system for pilot and large scale purification of biopharmaceutical products. The system performs equilibration, sample loading, elution, fraction collection, regeneration and washing according to the specified method. The method editor allows you to control the sequence of operations by time, liquid volume or column volume. The compact layout and multi-level arrangement of pipelines reduce the residual volume of the liquid path.
Models and versions
| Model | Consumption, l/min | Pipeline | Bubble trap, ml | Power, kW | Weight, kg |
|---|---|---|---|---|---|
| AP005D | 0.01-0.6 | DN08 | 80 | 1 | 160 |
| AP010S(D) | 0.02-1 | DN08 | 100 | 1 | 230 |
| AP030S(D) | 0.03-3 | DN10 | 300 | 1 | 230 |
| AP100S(D) | 0.3-8 | DN15 | 1500 | 2 | 300 |
| AP200S(D) | 0.5-16 | DN20 | 2200 | 2 | 400 |
| AP400S(D) | 1-40 | DN25 | 4000 | 4 | 520 |
| AP800S(D) | 2-80 | DN40 | 8000 | 5.5 | 600 |
| AP1600S(D) | 10-160 | DN50 | 16000 | 8.5 | 800 |
Product Features
Compact layout
Multi-level installation of pipelines reduces the occupied space and dead volume.Valve feedback
The opening and closing status of the valves is monitored in real time.Agile Methods
Automatic operations are set by time, volume and column volume.Own software
The software platform integrates method setup, operation execution and monitoring.Convenient control
The menu and graphical interface ensure sequential execution of operations.Remote monitoring
Remote monitoring of equipment condition is provided.Technical Parameters
| Parameter | Meaning |
|---|---|
| Flow Accuracy | ±2% |
| Maximum working pressure | 5 bar |
| Pressure sensor | -1...9 bar; accuracy ≤0.3% |
| pH range | 0-14; accuracy ±0.1 in pH range 2-12 |
| Electrical conductivity | 1 µS/cm - 300 mS/cm |
| UV detection | 200-400 or 200-800 nm; range 0-3 AU |
| Contact materials | 316L, PP, EPDM, borosilicate glass, PEEK |
| Operating temperature | 2-40 °C |
| Degree of protection | IP54 |
Frequently asked questions
What operations are performed automatically?
Equilibration, sample loading, elution, collection, regeneration and washing are performed according to the specified method.
How is the cleaning method set?
The sequence of operations can be specified by time, liquid volume, or column volume.
Is it possible to monitor the condition of the valves?
The system uses valve feedback and controls their opening and closing in real time.