UF/DF and filtration

Virus filtration system

Filtration system for a critical stage of the post-treatment process.

Product Description

Viral filtration system for the virus removal step in the purification of biopharmaceutical products. The three models differ in flow range and membrane area.

Control of pressure and electrical conductivity, constant pressure, flow and speed modes are provided. The filter module is selected taking into account the product and process conditions.

Models and versions

Models and versions: Virus filtration system
ModelConsumption, l/minMembrane area, m²Power, kWWeight, kg
VF01000.3-100.5-21.5250
VF02000.5-201-42260
VF04002.5-402-83300

Product Features

Three flow ranges

Versions VF0100, VF0200 and VF0400 for different capacities.

Process control

Measurement of pressure and electrical conductivity in the liquid path.

Membrane compatibility

Supported solutions include Millipore, Sartorius, Pall and Asahi Kasei.

Control Modes

Operation at constant pressure, flow or speed.

Technical Parameters

Technical Parameters: Virus Filtration System
ParameterMeaning
Maximum working pressure4 bar
Pressure sensor-1...9 bar; accuracy ≤0.3%
Electrical conductivity1 µS/cm - 300 mS/cm
Contact materials316L, PP, EPDM, PEEK
Operating temperature2-40 °C
Degree of protectionIP54
Power supply VF0100 / VF0200200-240 V AC, 50/60 Hz
Power supply VF04003AC, 340-420 V, 50/60 Hz

Frequently asked questions

How are the VF models different?

Flow range, membrane area, power and weight; values ​​are given in the table of versions.

Which membranes are compatible with the system?

Works with Millipore, Sartorius, Pall and Asahi Kasei filter modules.

What is the maximum working pressure?

The range has a maximum operating pressure of 4 bar.