UF/DF and filtration
Virus filtration system
Filtration system for a critical stage of the post-treatment process.
Product Description
Viral filtration system for the virus removal step in the purification of biopharmaceutical products. The three models differ in flow range and membrane area.
Control of pressure and electrical conductivity, constant pressure, flow and speed modes are provided. The filter module is selected taking into account the product and process conditions.
Models and versions
| Model | Consumption, l/min | Membrane area, m² | Power, kW | Weight, kg |
|---|---|---|---|---|
| VF0100 | 0.3-10 | 0.5-2 | 1.5 | 250 |
| VF0200 | 0.5-20 | 1-4 | 2 | 260 |
| VF0400 | 2.5-40 | 2-8 | 3 | 300 |
Product Features
Three flow ranges
Versions VF0100, VF0200 and VF0400 for different capacities.
Process control
Measurement of pressure and electrical conductivity in the liquid path.
Membrane compatibility
Supported solutions include Millipore, Sartorius, Pall and Asahi Kasei.
Control Modes
Operation at constant pressure, flow or speed.
Technical Parameters
| Parameter | Meaning |
|---|---|
| Maximum working pressure | 4 bar |
| Pressure sensor | -1...9 bar; accuracy ≤0.3% |
| Electrical conductivity | 1 µS/cm - 300 mS/cm |
| Contact materials | 316L, PP, EPDM, PEEK |
| Operating temperature | 2-40 °C |
| Degree of protection | IP54 |
| Power supply VF0100 / VF0200 | 200-240 V AC, 50/60 Hz |
| Power supply VF0400 | 3AC, 340-420 V, 50/60 Hz |
Frequently asked questions
How are the VF models different?
Flow range, membrane area, power and weight; values are given in the table of versions.
Which membranes are compatible with the system?
Works with Millipore, Sartorius, Pall and Asahi Kasei filter modules.
What is the maximum working pressure?
The range has a maximum operating pressure of 4 bar.